• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE CORPORATION GAMCATH; CATHETER, HEMODIALYSIS, NON-IMPLANTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE CORPORATION GAMCATH; CATHETER, HEMODIALYSIS, NON-IMPLANTED Back to Search Results
Catalog Number 112714
Device Problem Fluid/Blood Leak (1250)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/16/2021
Event Type  malfunction  
Manufacturer Narrative
The device was received for evaluation.A visual inspection observed a crack running in longitudinal direction, starting from the connector borderline and running along a grip.The crack was approximately 19-20 mm length.The lateral area near the thread showed some micro cracks.The reported condition was verified.The cause could not be determined.Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that during use with a gamcath, an external leak from the extension tube nearby a connector.There was no report of patient injury or medical intervention associated with this event.No additional information is available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GAMCATH
Type of Device
CATHETER, HEMODIALYSIS, NON-IMPLANTED
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
BAXTER HEALTHCARE - HECHINGEN CATHETER
linsenacker 1
ni
hechingen 72379
GM   72379
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 
2242702068
MDR Report Key11487351
MDR Text Key245324455
Report Number3004367028-2021-00001
Device Sequence Number1
Product Code MPB
Combination Product (y/n)Y
Reporter Country CodeJA
PMA/PMN Number
NI
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial
Report Date 03/15/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number112714
Device Lot NumberASKU
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/03/2021
Initial Date Manufacturer Received 02/18/2021
Initial Date FDA Received03/15/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-