SMITH & NEPHEW, INC. DYONICS POWERMINI WITH HAND CONTROLS; SAW, POWERED, AND ACCESSORIES
|
Back to Search Results |
|
Model Number 72201500 |
Device Problem
Unintended Power Up (1162)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 02/17/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
Internal complaint reference case: (b)(4).
|
|
Event Description
|
It was reported that during a shoulder scope surgery the powermini starts and stops on its own.The procedure was successfully completed without delay using a back-up device.No patient injury or other complications were reported.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
|
|
Manufacturer Narrative
|
The reported device, used in treatment, was received for evaluation.A visual inspection was performed on the product and no issue was observed.There was a relationship found between the returned device and the reported incident.Product failed functional testing with blade stall error and overheating.Cause of overheating and errors is a corroded motor/gearbox.The gearbox was found to be jammed.The complaint was confirmed and the root cause has been determined to be corrosion of the motor and gearbox assembly.A blade stall condition will result in increased current draw from the control unit which will heat the motor and hand piece housing.Factors which can contribute to gearbox corrosion include cleaning and sterilization methods and the chemicals involved.A review of the device history records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.A complaint history review concluded this was a repeat issue.
|
|
Search Alerts/Recalls
|
|
|