• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECTON, DICKINSON AND COMPANY BD INSYTE-N AUTOGUARD; CATHETER,INTRAVASCULAR,THERAPEUTIC,SHORT-TERM LESS THAN 30 DAYS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON, DICKINSON AND COMPANY BD INSYTE-N AUTOGUARD; CATHETER,INTRAVASCULAR,THERAPEUTIC,SHORT-TERM LESS THAN 30 DAYS Back to Search Results
Model Number 381511
Device Problems Material Fragmentation (1261); Difficult to Insert (1316); Activation, Positioning or Separation Problem (2906); Material Integrity Problem (2978); Material Split, Cut or Torn (4008); Physical Resistance/Sticking (4012)
Patient Problem Insufficient Information (4580)
Event Date 01/06/2021
Event Type  malfunction  
Event Description
Multiple complaints from nursing staff of bd intravascular catheters splintering or splitting.Upon single attempt, the plastic sheath of cahtheter splinters when needle is re-introduced.On initial insertion, there have been complaints of needle not cannulating the skin/vein easily and resistance is met.Iv infiltrates appear to be trending up.Below is some of the descriptions that have been shared: the catheter is hard to cannulate the skin "almost as the skin is tough and i know it is not"."i meet resistance if i need to reintroduce the needle into the catheter"."it doesn't seem to thread right"."it doesn't feel right when i puncture the skin"."when i pull it out the catheter is splintered or torn" (after iv attempt was unsuccessful).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD INSYTE-N AUTOGUARD
Type of Device
CATHETER,INTRAVASCULAR,THERAPEUTIC,SHORT-TERM LESS THAN 30 DAYS
Manufacturer (Section D)
BECTON, DICKINSON AND COMPANY
1 becton drive
franklin lakes NJ 07417
MDR Report Key11494550
MDR Text Key240081357
Report Number11494550
Device Sequence Number1
Product Code FOZ
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source User Facility
Reporter Occupation Risk Manager
Type of Report Initial
Report Date 03/10/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model Number381511
Device Catalogue Number381511
Device Lot Number0195143 0080379 0119474 026501
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? Yes
Date Report Sent to FDA03/10/2021
Event Location Hospital
Date Report to Manufacturer03/16/2021
Initial Date Manufacturer Received Not provided
Initial Date FDA Received03/16/2021
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Age1 DA
-
-