• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CAREFUSION 213, LLC 0113 CHLORAPREP ONE STEP HI LITE ORANGE; 2% W/V CHLORHEXIDINE GLUCONATE/70% V/V ISOPROPYL ALCOHOL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CAREFUSION 213, LLC 0113 CHLORAPREP ONE STEP HI LITE ORANGE; 2% W/V CHLORHEXIDINE GLUCONATE/70% V/V ISOPROPYL ALCOHOL Back to Search Results
Catalog Number 930815
Device Problems Defective Component (2292); Detachment of Device or Device Component (2907)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/11/2021
Event Type  malfunction  
Manufacturer Narrative
(b)(4) initial emdr submission.A follow up emdr will be submitted if additional information becomes available.(b)(4).Medwatch- (b)(4).
 
Event Description
Material no.: 930815.Batch no.: 0327867.It was reported the end of the handle came off causing the glass cartridge to fall out onto the floor.Per medwatch: describe the event or problem: the user of the device was preparing the choloraprep when glass cartridge shot out of the handle and shattered on the floor.No patient harm.
 
Manufacturer Narrative
The facility did provide not photos/samples to aid in our quality engineer¿s investigation.With the lack of sample provided, bd was unable to verify the failure mode as the end cap was detached from the applicator body.A device history record was reviewed, and no non-conformances was noted during the manufacturing of this lot.Although the complaint could not be verified, the root cause is attributed to the equipment station for the end cap placement unto the applicator body.Corrective actions were initiated which led to bd conducting a voluntary recall on certain lots of the chloraprep hi- lite orange 26 ml applicator.Bd has confirmed that some of the product, which included this lot, had an applicator end cap that was improperly secured during the manufacturing process which resulted in broken glass dropping out of the applicator.
 
Event Description
It was reported the end of the handle came off causing the glass cartridge to fall out onto the floor.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CHLORAPREP ONE STEP HI LITE ORANGE
Type of Device
2% W/V CHLORHEXIDINE GLUCONATE/70% V/V ISOPROPYL ALCOHOL
Manufacturer (Section D)
CAREFUSION 213, LLC 0113
1550 northwestern dr
el paso TX 79912
MDR Report Key11499917
MDR Text Key241960185
Report Number3004932373-2021-00134
Device Sequence Number1
Product Code KXG
Combination Product (y/n)Y
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type other
Type of Report Initial,Followup
Report Date 06/25/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/30/2023
Device Catalogue Number930815
Device Lot Number0327867
Initial Date Manufacturer Received 03/05/2021
Initial Date FDA Received03/16/2021
Supplement Dates Manufacturer Received06/25/2021
Supplement Dates FDA Received07/06/2021
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age33 YR
-
-