Model Number FOL0102
Device Problem Component Missing (2306)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/23/2021
Event Type  malfunction  
Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.
 
Event Description
It was reported that 2 pieces were missing in the master box.It was stated that instead of 50 pieces only 48 pieces were received.
 
Event Description
It was reported that 2 pieces were missing in the master box.It was stated that instead of 50 pieces only 48 pieces were received.
 
Manufacturer Narrative
The reported issue was inconclusive due to poor sample condition.Visual inspection noted two photo samples were received.Visual evaluation noted one photo shows a single statlock package, and the second photo showed the bulk packaging.Based on the photos received, the reported event was unable to be confirmed.A potential root cause for this reported failure was operator error.This device was not used for patient treatment or diagnosis.It was unknown whether the product had caused the reported failure.It was unknown whether the device met specifications or not.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.A labeling review was not performed because labeling could not have prevented the reported failure.H11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SL FOLEY SWIVEL SILICONE TRICOT 25BX
Type of Device
STATLOCK DEVICE
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
MDR Report Key11509684
MDR Text Key248820580
Report Number1018233-2021-01356
Device Sequence Number1
Product Code EYJ
UDI-Device Identifier00801741076114
UDI-Public(01)00801741076114
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,other
Type of Report Initial,Followup
Report Date 05/25/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date10/28/2023
Device Model NumberFOL0102
Device Catalogue NumberFOL0102
Device Lot NumberJUEX2169
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/23/2021
Initial Date Manufacturer Received 02/23/2021
Initial Date FDA Received03/17/2021
Supplement Dates Manufacturer Received05/25/2021
Supplement Dates FDA Received06/17/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Other;
-
-