Catalog Number 3035890011-3
Device Problem Unsealed Device Packaging (1444)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/10/2021
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Report source, foreign - event occurred in (b)(6).One picture has been received and the reported event was confirmed as it can be seen that the inner cement pouch sealing is damaged and that cement powder leaked outside the inner pouch.The product analysis can't be performed as the product was not returned.The complaint is related to a well-known event, previously evaluated and investigated.No further investigation is required as per 21 cfr part 820.198 complaint files §(b) and (c).Investigation results concluded that the reported event was due to packaging issue (sealing process).Corrective action has been initiated to address reported issue.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
It has been reported that, before the surgery, it was found (by zb staff) that the inner package had leaked.In addition it was reported that the product will not be returned due to the cost of shipment.No adverse event has been reported as a result of the malfunction.
 
Event Description
It has been reported that, before the surgery, it was found (by zb staff) that the inner package had leaked.In addition it was reported that the product will not be returned due to the cost of shipment.No adverse event has been reported as a result of the malfunction.
 
Manufacturer Narrative
(b)(4).This follow-up report is being submitted to relay additional information.This follow-up report is being submitted to relay corrected information.The device manufacturing quality record indicate that the released product met all requirements to perform as intended.Complaint history identified no additional complaints for the reported part/lot combination.A definitive root cause cannot be determined.Further invetigation is being conducted through our capa process.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIOMET BONE CEMENT R 40 -3
Type of Device
BONE CEMENT
Manufacturer (Section D)
BIOMET FRANCE S.A.R.L.
plateau de lautagne bp75
valence cedex 26903
FR  26903
Manufacturer (Section G)
BIOMET FRANCE S.A.R.L.
plateau de lautagne bp75
valence cedex 26903
FR   26903
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key11517176
MDR Text Key240781414
Report Number3006946279-2021-00038
Device Sequence Number1
Product Code LOD
Combination Product (y/n)N
Reporter Country CodeTH
PMA/PMN Number
N/A
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 04/11/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/18/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/31/2023
Device Catalogue Number3035890011-3
Device Lot Number826AAC0512
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received04/08/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/24/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-