Model Number CDVRA500Q
Device Problems Loose or Intermittent Connection (1371); Failure to Disconnect (2541)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/01/2021
Event Type  malfunction  
Event Description
It was reported that the patient presented for a system implant procedure.During the procedure, the physician was unable to engage the wrench inside the header.The physician was then unable to remove the wrench from the set screw and the set screw completely backed out of the header.The device was removed and replaced.The patient was in stable condition.
 
Manufacturer Narrative
The reported field event of a loose set screw was confirmed.Final analysis found septum material in the hex cavity which prevented insertion of the torque wrench.Once the material was removed, the set screw was able to be secured.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GALLANT VR
Type of Device
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
MDR Report Key11532719
MDR Text Key241515047
Report Number2017865-2021-12260
Device Sequence Number1
Product Code NIK
UDI-Device Identifier05415067031990
UDI-Public05415067031990
Combination Product (y/n)N
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 04/28/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/19/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/30/2022
Device Model NumberCDVRA500Q
Device Catalogue NumberCDVRA500Q
Device Lot NumberS000079070
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/16/2021
Was the Report Sent to FDA? No
Date Manufacturer Received04/09/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age54 YR
-
-