Model Number 71118200
Device Problem Break (1069)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/28/2021
Event Type  malfunction  
Manufacturer Narrative
Internal complaint reference number: (b)(4).
 
Event Description
It was reported that, during an unknown procedure, the flexible shaft w/cir connector was being used to ream the canal of a tibia and it sheared off.Instruments were outside the patient.There was no delay and procedure was completed with the same device.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
 
Manufacturer Narrative
H6: the device, intended for use in treatment, was not returned for evaluation.Therefore, product analysis could not be performed at this time.So the reported event could not be confirmed.A complaint history review found related failures; this failure mode will be monitored for future complaints for any necessary corrective actions.This device is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Based on this investigation, the need for corrective action is not indicated.Without the actual product involved and/or device information, our investigation cannot proceed.If the device or new information is received in the future, this complaint can be re-opened.No further investigation warranted for these complaints; however, we will continue to monitor for future complaints and investigate as necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLEXIBLE SHAFT W/CIR CONNECTOR
Type of Device
PROSTHESIS, SHOULDER, HEMI-, HUMERAL, METALLIC UNCEMENTED
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key11541603
MDR Text Key241445228
Report Number1020279-2021-02287
Device Sequence Number1
Product Code HSD
UDI-Device Identifier03596010436528
UDI-Public03596010436528
Combination Product (y/n)N
PMA/PMN Number
K121714
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 05/07/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/22/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number71118200
Device Catalogue Number71118200
Device Lot NumberUNKNOWN
Date Manufacturer Received05/04/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-