Model Number NSLX137C |
Device Problems
Activation, Positioning or Separation Problem (2906); Difficult to Open or Close (2921); No Apparent Adverse Event (3189)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/01/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).Batch #: unknown.As the device was not returned, an analysis investigation could not be performed.A conclusion could not be reached as to what may have caused or contributed to the event.Complaint information is trended on a regular basis to determine if further investigation is warranted.The lot/ batch was not provided; therefore, the manufacturing records evaluation could not be performed.
|
|
Event Description
|
It was reported that during an esophagus procedure, the device worked really well at the beginning.No issues.However, the sticking started occurring after about 90 minutes.The non-fatty tissue was not easy to coag at this point verses the beginning.The device was cleaned when there was a lot of tissue.It looked like it was lacking the silicone coating, so heavy sticking occurred.The device stuck on a 6mm artery and the surgeon needed to separate the device with force.Bleeding occurred and it was clipped to address it.There were no patient consequences.
|
|
Manufacturer Narrative
|
(b)(4).Date sent: 4/1/2021.Upon review of the information provided, it was concluded that this event does not meet the fda defined criteria for a reportable event and is being considered not reportable.
|
|