• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON ENDO-SURGERY, LLC. ENSEAL X1 CURVED 37CM SHAFT; ELECTROSURGICAL CUTTING & COAGULATION & ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON ENDO-SURGERY, LLC. ENSEAL X1 CURVED 37CM SHAFT; ELECTROSURGICAL CUTTING & COAGULATION & ACCESSORIES Back to Search Results
Model Number NSLX137C
Device Problems Activation, Positioning or Separation Problem (2906); Difficult to Open or Close (2921); No Apparent Adverse Event (3189)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/01/2021
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Batch #: unknown.As the device was not returned, an analysis investigation could not be performed.A conclusion could not be reached as to what may have caused or contributed to the event.Complaint information is trended on a regular basis to determine if further investigation is warranted.The lot/ batch was not provided; therefore, the manufacturing records evaluation could not be performed.
 
Event Description
It was reported that during an esophagus procedure, the device worked really well at the beginning.No issues.However, the sticking started occurring after about 90 minutes.The non-fatty tissue was not easy to coag at this point verses the beginning.The device was cleaned when there was a lot of tissue.It looked like it was lacking the silicone coating, so heavy sticking occurred.The device stuck on a 6mm artery and the surgeon needed to separate the device with force.Bleeding occurred and it was clipped to address it.There were no patient consequences.
 
Manufacturer Narrative
(b)(4).Date sent: 4/1/2021.Upon review of the information provided, it was concluded that this event does not meet the fda defined criteria for a reportable event and is being considered not reportable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENSEAL X1 CURVED 37CM SHAFT
Type of Device
ELECTROSURGICAL CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
MDR Report Key11545916
MDR Text Key241799416
Report Number3005075853-2021-01531
Device Sequence Number1
Product Code GEI
UDI-Device Identifier10705036015468
UDI-Public10705036015468
Combination Product (y/n)N
PMA/PMN Number
K172580
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,study
Type of Report Initial,Followup
Report Date 03/01/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/22/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberNSLX137C
Device Catalogue NumberNSLX137C
Was Device Available for Evaluation? No
Date Manufacturer Received03/31/2021
Patient Sequence Number1
Treatment
GENERATOR
-
-