Model Number CI-1600-04 |
Device Problems
Expulsion (2933); Material Protrusion/Extrusion (2979); Adverse Event Without Identified Device or Use Problem (2993)
|
Patient Problems
Erosion (1750); Pain (1994); Ulcer (2274); Tissue Breakdown (2681)
|
Event Date 08/20/2021 |
Event Type
Injury
|
Event Description
|
The recipient reportedly experienced a sore at the implant site.The recipient presented with skin flap breakdown and device extrusion.The recipient ceased device use.The recipient's device was explanted and the skin flap was closed.The recipient will be reimplanted once healed.
|
|
Manufacturer Narrative
|
The recipient has reportedly completely healed.The recipient's device was not explanted.Explant surgery is scheduled.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
|
Manufacturer Narrative
|
The recipient's device was explanted.The recipient will be reimplanted at a later date.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
|
Manufacturer Narrative
|
The recipient's device was explanted.The recipient was re-implanted with another advanced bionics cochlear device.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
|
Manufacturer Narrative
|
The recipient's device was reportedly not explanted.Revision surgery is scheduled.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
|
Manufacturer Narrative
|
Advanced bionics considers the investigation into this reportable event as closed.The external visual inspection revealed that the electrode was severed.This is believed to have occurred during revision surgery.The device passed the photographic imaging inspection.System lock was verified.The electrode condition prevented an electrical test from being performed.The device passed the electrical and mechanical tests performed.This device was explanted for medical reasons.The device passed the tests performed.This version of the hires ultra device is no longer distributed.This is the final report.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
|
Manufacturer Narrative
|
Advanced bionics considers the investigation into this reportable event as closed.The recipient reportedly has not had any post operative complications.No further treatment details will be provided.The external visual inspection revealed that the electrode was severed.This is believed to have occurred during revision surgery.The device passed the photographic imaging inspection.System lock was verified.The electrode condition prevented an electrical test from being performed.The device passed the electrical and mechanical tests performed.This device was explanted for medical reasons.The device passed the tests performed.This version of the hires ultra device is no longer distributed.This is the final report.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
|