MAUDE Adverse Event Report: OLYMPUS WINTER & IBE GMBH HF-RESECTION ELECTRODE, LOOP, 24 FR., 0.35 WIRE, 12, STERILE, SINGLE USE, 12 PC; HF-RESECTION ELECTRODES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS WINTER & IBE GMBH HF-RESECTION ELECTRODE, LOOP, 24 FR., 0.35 WIRE, 12, STERILE, SINGLE USE, 12 PC; HF-RESECTION ELECTRODES Back to Search Results
Model Number A22201C
Device Problems Break (1069); Fracture (1260); Material Fragmentation (1261); Mechanical Problem (1384)
Patient Problem Foreign Body In Patient (2687)
Event Date 03/18/2021
Event Type  malfunction  
Manufacturer Narrative
The suspect medical device has not yet been returned to olympus for evaluation/investigation.Therefore, the exact cause of the user's experience and the reported phenomenon could not be determined and is being judged as unknown.However, if the suspect medical device is returned for evaluation/investigation or additional significant information becomes available, this report will be updated.
 
Event Description
Olympus was informed that during a therapeutic transurethral resection of a bladder tumor (turbt) procedure, the loop wire at the distal end of the hf resection electrode broke and fragments fell into the patient.It is unknown whether or not fragments may have remained inside the patient.An x-ray examination was to be performed to confirm that no fragments remained inside the patient, the results of which have not been provided.However, the intended procedure was completed using a similar device and there was no report about an adverse event or patient injury.
 
Manufacturer Narrative
The suspect medical device was returned to the manufacturer for investigation.The investigation confirmed that the loop wire at the distal end of the resection electrode is broken and the loop burnt down subsequently.During the announced x-ray examination no parts were found inside the patient.The cause of this event was attributed to wear and tear during use.A material or quality problem can be excluded since a manufacturing and quality control review was performed for the affected lot number of the hf resection electrode without showing any abnormalities.The case will be closed on olympus side with no further actions.However, the reported phenomenon will be recorded for trending and surveillance purposes and the user will be informed about the investigation results.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HF-RESECTION ELECTRODE, LOOP, 24 FR., 0.35 WIRE, 12, STERILE, SINGLE USE, 12 PC
Type of Device
HF-RESECTION ELECTRODES
Manufacturer (Section D)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg 22045
GM  22045
Manufacturer (Section G)
OLYMPUS MEDICAL PRODUCTS CZECH, SPOL. S R.O.
telickova 457/29
prerov
Manufacturer Contact
daniel wladow
kuehnstrasse 61
hamburg 22045
GM   22045
4940669662
MDR Report Key11552490
MDR Text Key244922299
Report Number9610773-2021-00103
Device Sequence Number1
Product Code FAS
UDI-Device Identifier14042761036648
UDI-Public14042761036648
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
K931763
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 06/01/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberA22201C
Device Catalogue NumberA22201C
Device Lot Number1000059075
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/17/2021
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 03/18/2021
Initial Date FDA Received03/23/2021
Supplement Dates Manufacturer Received05/27/2021
Supplement Dates FDA Received06/01/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/27/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-