On (b)(6) 2021, a 25 mm amplatzer amulet left atrial appendage occluder was selected for implant.During the procedure, the device was prepared and advanced through the delivery sheath into the left atrial appendage.After two attempts, the device was fully recaptured into the delivery sheath in order to reposition the sheath into the left atrial appendage (laa).Resistance was noted resistance when trying to redeploy the device and it was decided to fully retrieve the device into the loader.Upon inspection, a thrombus was found on the device and the cable connected to the device.The physician suspected a clotting disorder as the cause.A larger heparin dose was administered and a new 25 mm amplatzer amulet left atrial appendage occluder and delivery system were used to complete the procedure.There was no clinically significant delay in the procedure and the patient remained hemodynamically stable throughout.The patient is stable.
|
The reported event of thrombus on the device and cable could not be confirmed.A more comprehensive assessment could not be performed as the device was not returned for analysis.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all specifications.Based on the information received, the cause of the reported incident could not be conclusively determined.Per the instructions for use (ifu) of the delivery sheath, artmt600034452 rev.B, "when placing a device using an amplatzer¿ amulet¿ delivery sheath, refer to the instructions for use provided with the device." the amulet ifu, artmt600034453 rev.C, states "if the device is retracted farther than the radiopaque markers (fully recaptured), the device and the sheath must both be removed and replaced.Failure to replace both the device and the sheath may result in sheath and/or device malfunction.".
|