BIOSENSE WEBSTER INC THERMOCOOL SMART TOUCH ELECTROPHYSIOLOGY CATHETER; SIMILAR DEVICE D133601, PMA # P030031/S053
|
Back to Search Results |
|
Catalog Number D133605IL |
Device Problem
Positioning Problem (3009)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/09/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
The bwi product analysis lab received the device for evaluation on 2/1/2021.The device evaluation was completed on 2/26/2021.The returned device was visually inspected and it was found with reddish material and a hole in the pebax with internal parts exposed.The deflection curve was tested and met the specification.A manufacturing record evaluation was performed for the finished device, and no internal action related to the complaint was found during the review.The deflection issue reported by the customer was not confirmed.The root cause of the pebax broken cannot be related to the manufacturing process since there is evidence that the device was manufactured in accordance with documented specification and procedures.It could be related to the handling of the device during the procedure; however, this cannot be conclusively determined.The device was working correctly.If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.Manufacturer's reference number: (b)(4).
|
|
Event Description
|
It was reported that a patient underwent an atrial fibrillation (afib) procedure with a thermocool® smart touch¿ electrophysiology catheter and the biosense webster, inc.Product analysis lab observed a hole on the pebax with internal parts exposed.Initially it was reported that there was a deflection issue.During the procedure, the catheter was unable to deflect or relax completely.A second catheter was used to complete the procedure.There was no patient consequence reported.The deflection issue was assessed as not mdr reportable.The most likely consequence was an intraprocedural delay.The potential risk that it could cause or contribute to a serious injury or death was remote.The biosense webster, inc.Product analysis lab received the device for evaluation and observed on 2/23/2021 reddish material inside the pebax and a hole on it with internal parts exposed.The returned condition of the hole on the pebax with internal parts exposed was assessed as mdr reportable.The awareness date is 2/23/2021.
|
|
Search Alerts/Recalls
|
|
|