Additional information: d9, g3, h2, h3, h6 the distal 5¿ of a decapolar, inquiry steerable diagnostic catheter shaft was received for evaluation.The rest of the catheter was not received.The cause of the deflection issue was unable to be confirmed due to the condition of the returned device.The catheter shaft was fractured and detached and electrodes 2 and 5-10 were displaced.Functional testing was not possible due to the aforementioned damage.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.The cause of the deflection issue remains unknown.The cause of the damage, displaced electrodes, and withdrawal difficulty are consistent with damage during use.
|