According to the reporter, during pylorus gastrectomy, when using a clip on blood vessel, slight dissection was performed with the tip of the clip.When the clip was loaded, the jaws were out of alignment and the parts inside were exposed.A new clip was opened and used to complete the case.There was no patient injury.
|
Evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted the driver of the component was found to be broken.The function of the driver is to advance through the jaw channel and make them close.When the driver moves backwards, jaw will open.When the driver is broken, the clips will not load properly into the jaws as expected.It was reported that the jaws were misaligned.The reported issue was confirmed.A component failure was identified during product analysis.Internal process improvements have been initiated to mitigate this issue.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|