Model Number CDHFA500Q |
Device Problems
High impedance (1291); Difficult to Insert (1316); Defective Device (2588)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 03/04/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.
|
|
Event Description
|
It was reported that the patient presented for an implant procedure.During the procedure, it was noted that the right ventricular pacing impedance was high.The physician disconnected and reconnected the lead multiple times until appropriate impedance was obtained.The physician found that it was difficult to insert the lead into the header.The procedure was completed successfully.The patient was in stable condition.
|
|
Manufacturer Narrative
|
Correction: medical device problem code should have been difficult to insert rather than defective device.
|
|