• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. EVOS SMALL 2.7MM/3.5MM SCREW CADDY LID; TRAY, SURGICAL, INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. EVOS SMALL 2.7MM/3.5MM SCREW CADDY LID; TRAY, SURGICAL, INSTRUMENT Back to Search Results
Model Number 71170225
Device Problem Flaked (1246)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/28/2021
Event Type  malfunction  
Event Description
It was reported that, during set up or inspection, the evos small 2.7mm/3.5mm screw caddy lid has pitting and chipping that appears to be dropping flakes onto the screws.Incident occurred before the procedure; therefore, there was no patient involvement.
 
Manufacturer Narrative
The device, intended for use in treatment, was not returned for evaluation.Therefore, product analysis could not be performed at this time.So the reported event could not be confirmed.A complaint history review found related failures; this failure mode will be monitored for future complaints for any necessary corrective actions.This device is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Based on this investigation, the need for corrective action is not indicated.Without the actual product involved and/or device information, our investigation cannot proceed.If the device or new information is received in the future, this complaint can be re-opened.No further investigation warranted for these complaints; however, we will continue to monitor for future complaints and investigate as necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVOS SMALL 2.7MM/3.5MM SCREW CADDY LID
Type of Device
TRAY, SURGICAL, INSTRUMENT
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key11566372
MDR Text Key242141520
Report Number1020279-2021-02334
Device Sequence Number1
Product Code FSM
UDI-Device Identifier00885556636152
UDI-Public00885556636152
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 05/06/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number71170225
Device Catalogue Number71170225
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 03/01/2021
Initial Date FDA Received03/24/2021
Supplement Dates Manufacturer Received04/27/2021
Supplement Dates FDA Received05/06/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-