NSK NAKANISHI INC FORZA M5 1:5 ATTACHMENT SHORTY FG HIGH SPEED OPTIC; HANDPIECE Back to Search Results
Model Number 5021285U0
Device Problems Break (1069); Overheating of Device (1437)
Patient Problem Superficial (First Degree) Burn (2685)
Event Date 03/09/2021
Event Type  malfunction  
Event Description
Customer reports two broken, overheating and bad sounding handpieces.A patient's lower right lip and lip inside cheek was burned (b)(6), male) during a crown prep.Topical [benzocaine] was applied and triple antibiotic ointment and 800mg ibuprofen was advised.The doctor does not think there will be scarring.In the same report, the customer said another patient's lip was burned a few weeks prior (no pt information was provided).The doctor says he can't always know when they are overheating and if the patient's lip is numb, they can't feel it either.Plenty of water is run through the handpieces, fresh burs are used and the handpieces are run through an oil machine between each use.
 
Event Description
Customer reports two broken, overheating and bad sounding handpieces.A patient's lower right lip and lip inside cheek was burned (initials he, age 18, male) during a crown prep.Topical [benzocaine] was applied and triple antibiotic ointment and 800mg ibuprofen was advised.The doctor does not think there will be scarring.In the same report, the customer said another patient's lip was burned a few weeks prior (no pt information was provided).The doctor says he can't always know when they are overheating and if the patient's lip is numb, they can't feel it either.Plenty of water is run through the handpieces, fresh burs are used and the handpieces are run through an oil machine between each use.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FORZA M5 1:5 ATTACHMENT SHORTY FG HIGH SPEED OPTIC
Type of Device
HANDPIECE
Manufacturer (Section D)
NSK NAKANISHI INC
700 shimohinata kanuma tochigi
322-8666
JA 
MDR Report Key11576576
MDR Text Key243902461
Report Number1032227-2021-00003
Device Sequence Number1
Product Code EGS
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Distributor
Type of Report Initial,Followup
Report Date 03/25/2021,03/29/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number5021285U0
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/25/2021
Was the Report Sent to FDA? Yes
Date Report Sent to FDA03/25/2021
Distributor Facility Aware Date03/12/2021
Event Location Other
Date Report to Manufacturer03/26/2021
Initial Date Manufacturer Received Not provided
Initial Date FDA Received03/26/2021
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received03/29/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age18 YR
-
-