Model Number 397002-001 |
Device Problems
Battery Problem (2885); Audible Prompt/Feedback Problem (4020)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 03/22/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
The companion 2 driver will be evaluated by syncardia.The results will be provided in a follow-up mdr.
|
|
Event Description
|
While performing a routine evaluation, a syncardia technician reported that the companion 2 driver exhibited an emergency battery error alarm.
|
|
Manufacturer Narrative
|
Communication with the emergency battery was established and the smbus data was retrieved and reviewed.The battery evaluation observed no permanent faults that would indicate a malfunction of the emergency battery.The emergency battery was then subjected to a ten-minute battery discharge test.After 7 minutes the emergency battery became depleted and the driver's main buzzer alarmed.The emergency battery error alarm was then displayed.This confirmed the customer-reported issue.The root cause was determined to be a malfunction of the emergency battery.This issue will be monitored and trended as part of the customer experience process.Syncardia has completed its investigation and is closing this file.Ce 5374 and ce 5504 follow-up report 1.
|
|