SMITH & NEPHEW, INC. GII MIS DCF AP CT BLK 6; PROSTH, KNEE, PATE/FEMOROTIBIAL, SEMI-CONST, UNCEM, PORO, COAT, POLY/METAL/POLY
|
Back to Search Results |
|
Model Number 71441154 |
Device Problem
Material Deformation (2976)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 03/04/2021 |
Event Type
malfunction
|
Event Description
|
It was reported that, component are unfit for use.The gii mis dcf ap ct blk 3 have burrs.No case reported; therefore, there was no patient involvement.
|
|
Manufacturer Narrative
|
H3, h6: the associated device, intended for use in treatment, was returned and evaluated.A visual inspection of the returned device confirmed the stated failure mode.The device has several burrs, nicks, and scratches rendering the device inoperative.The device shows signs of extensive use.A review of complaint history did not reveal additional complaints for the listed batch.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we do not have reason to suspect that the product failed to meet any product specifications at the time of manufacture.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
|
|
Search Alerts/Recalls
|
|
|