SMITH & NEPHEW, INC. GII OVAL PATELLAR PEG DRILL; PROSTH, KNEE, PATE/FEMOROTIBIAL, SEMI-CONST, UNCEM, PORO, COAT, POLY/METAL/POLY
|
Back to Search Results |
|
Model Number 71440743 |
Device Problems
Break (1069); Dull, Blunt (2407); Naturally Worn (2988)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 03/04/2021 |
Event Type
Injury
|
Event Description
|
It was reported that, during a knee revision surgery it was noticed that the gii oval patellar peg drill was worn, dull and broken.The procedure was completed without delay using a smith and nephew back-up device.No other complications were reported.
|
|
Manufacturer Narrative
|
H3, h6: the device, used in treatment, was returned for evaluation a visual inspection confirms the peg drill is dull and worn.The device shows significant signs of wear/usage.A complaint history review found related failures; this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the manufacturing records did not reveal any manufacturing or material abnormalities that could have caused or contributed to the reported incident.A review of the risk management file revealed this failure mode was previously identified.This device is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Expected wear/tear is likely the probable cause of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaints will be reopened.No further investigation is warranted for these complaints; however we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
|
|
Search Alerts/Recalls
|
|
|