As the lot number for the device was provided, a review of the device history records will be performed.The sample was not returned to the manufacturer for inspection/evaluation.Therefore, the investigation of the reported event is inconclusive.Based upon the available information, the definitive root cause for this event is unknown.The instructions for use (ifu) is adequate for the reported device/patient code(s) and provides general instructions for use, as well as warnings, precautions and potential complications associated with the device.Upon receipt of new or additional information, a follow-up report will be submitted as applicable.(expiry date: 07/2020).
|
Manufacturing review: a review of manufacturing records was not performed, as additional complaints have not been reported for this lot.Based on the information available it is not reasonably suggested that a manufacturing process may have caused or contributed to the reported issue.However, the lot history records of this lot were reviewed with special attention to the manufacturing and inspection of this product and the product was found to have met the specification prior to shipment.Investigation summary: the device was not available, images have not been provided.Based on review of the lot history records and the labeling of the device the expiry date of the sample was overdue when it was used on patient.Expiry date: 13.Jul 2020 versus date of event: 27.Feb 2021.Information confirming the event e.G., image from cathlab was not provided which leads to inconclusive evaluation result.Based on the information available the investigation is closed with inconclusive result.A definite root cause for the reported event could not be determined.Labeling review: in reviewing the relevant labeling for this product, the potential issue was found addressed.The instruction for use state: 'do not use the device after the ¿use by¿ date specified on the label.' h10: the catalog number identified in section d4 has not been cleared in the us but is similar to the venovo venous stent products that are cleared in the us.The pro code and 510 k number for the venovo venous stent products are identified in d2 and g4.H10: d4 (expiry date: 07/2020), g3 h11: h6 (method and conclusion) h11: section a through f ¿ the information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : device not returned.
|