U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DRÄGERWERK AG & CO. KGAA PERSEUS A500; ANESTHESIA UNITS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DRÄGERWERK AG & CO. KGAA PERSEUS A500; ANESTHESIA UNITS Back to Search Results
Model Number MK06000
Device Problems Gas Output Problem (1266); Inappropriate or Unexpected Reset (2959); Protective Measures Problem (3015); Reset Problem (3019)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/11/2021
Event Type  malfunction  
Manufacturer Narrative
The investigation was just started.The results will be provided in a follow-up report.
 
Event Description
It was reported that the device was in use in a surgery, when the curves started to go slower and slower, the user could see more pulses/ventilation periods on the screen than normal.The users was just about to change picture/view on the device when suddenly the screen went grey.According to the user the ventilation also stopped.The user was just about to start manual ventilation, when the screen came back on(user estimates around 10-20 seconds) and ventilation started again.No injury reported.
 
Manufacturer Narrative
Based on the logfile analysis, an already known vulnerability of the perseus medibus interface was determined to be the root cause of the reboot.The reboot was obviously caused by large medibus incompatible data packages leading to a processor overload and finally a reboot.Only sw versions 2.0n are affected by the identified medibus vulnerability.The reboot results in a short-term outage of therapy functions for not more than 15 seconds accompanied by a corresponding alarm, before the therapy is resumed with the last valid settings.As also reported, the ventilation could be resumed after the reboot without further problems.The software sub module for the medibus interface has already been improved and will be distributed as part of the new perseus software release 2.03.The distribution of the sw2.03 will be carried out with a field safety corrective action.All customers of draeger perseus a500 anesthesia machines equipped with sw2.0n were informed with a field safety notice in (b)(6) 2020 about the identified risk.All devices equipped with the affected sw versions 2.0n will be upgraded to sw2.03.
 
Event Description
It was reported that the device was in use in a surgery, when the curves started to go slower and slower, the user could see more pulses/ventilation periods on the screen than normal.The users was just about to change picture/view on the device when suddenly the screen went grey.According to the user the ventilation also stopped.The user was just about to start manual ventilation, when the screen came back on(user estimates around 10-20 seconds) and ventilation started again.No injury reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PERSEUS A500
Type of Device
ANESTHESIA UNITS
Manufacturer (Section D)
DRÄGERWERK AG & CO. KGAA
moislinger allee 53-55
lübeck 23542
GM  23542
MDR Report Key11587692
MDR Text Key244182236
Report Number9611500-2021-00136
Device Sequence Number1
Product Code BSZ
UDI-Device Identifier04048675253600
UDI-Public(01)04048675253600(11)190923(93)MK06000-37
Combination Product (y/n)N
PMA/PMN Number
K133886
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Type of Report Initial,Followup
Report Date 05/10/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMK06000
Device Catalogue NumberMK06000
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 03/12/2021
Initial Date FDA Received03/30/2021
Supplement Dates Manufacturer Received05/03/2021
Supplement Dates FDA Received05/11/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
NA.; NA.
-
-