Model Number 436-28-010
Device Problems Device Dislodged or Dislocated (2923); Appropriate Term/Code Not Available (3191)
Patient Problem Insufficient Information (4580)
Event Date 03/01/2021
Event Type  Injury  
Manufacturer Narrative
Additional reporting on this event will be provided as a supplemental report to this document if it becomes available.
 
Event Description
Second revision surgery - patient dislocated after a hip revision.A proximal femur was the component used.
 
Manufacturer Narrative
Manufacturer narrative: the reason for this revision surgery was reported as dislocation.The previous surgery and the surgery detailed in this event occurred 12 days apart.Initial or prolonged hospitalization was required.The healthcare professional indicated there was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The devices were disposed of at hospital and not made available to djo surgical for examination.A review of the device history records (dhr) show that the reported components used in the previous surgery, when released for use, met design and manufacturing requirements.There were no non-conforming material reports (ncmr) associated with the products that may have contributed to the reported event.The devices were verified to have gone through an acceptable sterilization process and were within its expiration date at the time of the previous surgery.Customer complaint history of the reported devices showed no present trends or on-going issues that are needing a review.The root cause of this compliant was a revision surgery due to dislocation.There were no findings during this evaluation that indicate the reported devices were defective.No information was submitted with the complaint regarding pre-existing conditions of the patient or any activities the patient was involved in that may have contributed to the event.Due to short time between previous and revision surgery, it is possible that the event may have occurred due to lack of post-operative care, patient noncompliance with medical instructions, improper surgical technique, patient activities or trauma.There are multiple factors that may also contribute to an event that are outside the control of djo surgical.There are no indications of a product or process issue affecting implant safety or effectiveness.Additional reporting on this event will be provided as a supplemental report to this document if it becomes available.
 
Event Description
Revision surgery - patient dislocated after a hip revision.A proximal femur was the component used.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FMP HIP
Type of Device
LINER, ACETABULAR, FMP CONSTRAINED 28XMP10
Manufacturer (Section D)
ENCORE MEDICAL L.P.
9800 metric blvd.
austin TX 78758 5445
MDR Report Key11590546
MDR Text Key242988701
Report Number1644408-2021-00231
Device Sequence Number1
Product Code KWZ
UDI-Device Identifier00888912085755
UDI-Public(01)00888912085755
Combination Product (y/n)N
PMA/PMN Number
K023794
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Type of Report Initial,Followup
Report Date 05/20/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date05/20/2021
Device Model Number436-28-010
Device Catalogue Number436-28-010
Device Lot Number502C1008
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 03/01/2021
Initial Date FDA Received03/30/2021
Supplement Dates Manufacturer Received04/22/2021
Supplement Dates FDA Received05/20/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
436-28-000 LOT 496C1088.; 436-28-000 LOT 496C1088
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age66 YR
-
-