• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL ADVISOR HD GRID MAPPING CATHETER, SENSOR ENABLED; CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL ADVISOR HD GRID MAPPING CATHETER, SENSOR ENABLED; CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING Back to Search Results
Model Number D-AVHD-DF16
Device Problem Entrapment of Device (1212)
Patient Problem Unspecified Tissue Injury (4559)
Event Date 03/18/2021
Event Type  Injury  
Manufacturer Narrative
The results of the investigation are inconclusive since the device was not returned for analysis.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.Based on the information received, the cause of the reported incident could not be conclusively determined.
 
Event Description
During an atrial tachycardia (at) ablation procedure, after successful ablation of atrial tachycardia and the patient was in a sinus rhythm, the advisor hd grid was advanced for mapping the sinus rhythm to confirm the location of the sinus node.After mapping, the device was attempted to be removed, however there was resistance felt.An x-ray revealed the hd grid was entangled with two other diagnostic catheters near the femoral iliac vein.The devices were attempted to be removed with no success.Vascular surgery was called for consult, and a snare requiring an additional subclavian access was used to successfully used to untangle and removed the devices.The patient is currently in stable condition.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADVISOR HD GRID MAPPING CATHETER, SENSOR ENABLED
Type of Device
CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING
Manufacturer (Section D)
ST. JUDE MEDICAL
5050 nathan lane north
plymouth MN 55442
Manufacturer (Section G)
ST. JUDE MEDICAL
5050 nathan lane north
plymouth MN 55442
Manufacturer Contact
stephanie o' sullivan
5050 nathan lane north
plymouth, MN 55442
6517565400
MDR Report Key11594985
MDR Text Key243126770
Report Number3005334138-2021-00193
Device Sequence Number1
Product Code DRF
UDI-Device Identifier05415067028198
UDI-Public05415067028198
Combination Product (y/n)N
Reporter Country CodeHK
PMA/PMN Number
K172393
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial
Report Date 03/31/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date10/31/2023
Device Model NumberD-AVHD-DF16
Device Catalogue NumberD-AVHD-DF16
Device Lot Number7731787
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 03/18/2021
Initial Date FDA Received03/31/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/09/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-