MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SYLMAR) MERLIN PROGRAMMER; PROGRAMMER, PACEMAKER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SYLMAR) MERLIN PROGRAMMER; PROGRAMMER, PACEMAKER Back to Search Results
Model Number 3650
Device Problems Program or Algorithm Execution Problem (4033); Program or Algorithm Execution Failure (4036)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/08/2021
Event Type  malfunction  
Manufacturer Narrative
The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.Further information was requested but not received.
 
Event Description
During capture threshold testing, the user released the button on the programmer to terminate the testing as expected, however, the implantable cardioverter defibrillator continued to decrement.The programmer was shut off prior to loss of capture for the dependent patient.The patient was in stable condition.
 
Manufacturer Narrative
The reported event of loss of capture during threshold testing was confirmed.The cause of the event was attributed to the programmer's failure to execute the algorithm to terminate testing.No other anomalies were found.
 
Event Description
Decrement test field action issued by abbott on 10 march 2022.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MERLIN PROGRAMMER
Type of Device
PROGRAMMER, PACEMAKER
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
sonali arangil
15900 valley view court
sylmar, CA 91342
MDR Report Key11597046
MDR Text Key243201421
Report Number2017865-2021-13398
Device Sequence Number1
Product Code NIK
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Non-Healthcare Professional
Remedial Action Recall
Type of Report Initial,Followup,Followup,Followup,Followup
Report Date 04/06/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number3650
Device Catalogue Number3650
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 03/08/2021
Initial Date FDA Received03/31/2021
Supplement Dates Manufacturer Received04/05/2021
06/29/2021
09/24/2021
03/10/2022
Supplement Dates FDA Received04/12/2021
06/29/2021
09/24/2021
04/07/2022
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction Number2017865-03/10/22¿001¿C
Patient Sequence Number1
Patient Age46 YR
Patient SexMale
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-