Updated fields: b4, b6, b7, d4 (version or model#), d11, g4, g7, h2, h4, h6 (type of investigation), h10, h11.Corrected field: a1, b5, e1 (event site email), e2, e3, g1 (contact person ¿ mfg site), g3.Analysis of production: (3331) the device history record review concluded that there were no ncmrs, rework, or deviations documented for the reported lot/serial number.Based on the dhr/lhr review results, it was determined that there is no relation between the manufacturing process and the reported failure.Historical data analysis: (4109) the review of the historical data indicates that no other similar complaint was reported for the same lot/serial number and reported failure mode.Trend analysis: (4110) the overall 24 month product complaint trend data for the period apr 2019 through mar 2021 was reviewed.There were no triggers identified for the review period.
|