Device evaluation found output error, testing failed the hand switch test lh (left handpiece) and rh (right handpiece) # 9 due to faulty auxiliary board.The output pbdes 200w testing was observed to be too low due to faulty psu (power supply unit) board.In addition, the unit was found running an old software version and needs to be upgraded.Unrelated to the reported issue, the unit was found missing four (4) mounting feet and several scratches were observed on the front panel and top case.Review of fault log found error 100 ref 18, one time (1) indicated continuous test timing.Cpu checks the timing completed within a certain time frame, 600 ref 14, five times (5) indicated foot switch mode pedal stuck.The usual cause is that the relevant foot pedal is held down by the user or there is a faulty foot pedal.100 ref 10, one time (1) indicated software execution failure.The generator software routines were interrupted unexpectedly by a watchdog reset.This can be caused by power ¿brown outs¿ or internal faults and 400 ref 12, three times (3) indicated footswitch mode pedal stuck.The usual cause is that the relevant foot pedal is held down during power on self-test or there is a faulty foot pedal.Based on evaluation findings, the reported issue was identified to be attributed to a faulty auxiliary board and psu attributed to electronic component failure.This report will be supplemented accordingly following investigations.
|