Model Number 6660 |
Device Problem
Wireless Communication Problem (3283)
|
Patient Problem
Inadequate Pain Relief (2388)
|
Event Date 03/12/2021 |
Event Type
Injury
|
Event Description
|
It was reported that the patient's ipg was inoperable following a patient procedure.Reportedly, the ipg was placed into surgery mode prior to the procedure, but was unresponsive following the procedure.As a result, the ipg was explanted and replaced.Surgical intervention resolved the issue.
|
|
Manufacturer Narrative
|
Correction, product information has been updated.The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.
|
|
Manufacturer Narrative
|
The device is included in the neuromodulation implantable pulse generator (ipg) inoperable when exposed to monopolar electrosurgery advisory notice issued by abbott on 02 june 2017.The reported event for inoperable ipg was confirmed.It was determined the device was running the service application software.This condition is consistent with electrocautery use during surgery.Per event details, the ipg no longer communicated following a lead revision procedure.There is guidance in the clinicians manual regarding proper handling of the device when using electrocautery.As a result of this finding, actions have been taken to prevent reoccurrence.
|
|
Search Alerts/Recalls
|