Model Number N/A |
Device Problem
Gel Leak (1267)
|
Patient Problems
Abrasion (1689); No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 03/17/2021 |
Event Type
Injury
|
Manufacturer Narrative
|
This event has been recorded by zimmer biomet under (b)(4).Once the investigation is completed, a follow-up/final report will be submitted.
|
|
Event Description
|
It was reported that during surgery air was heard to be exiting from head of hand piece.No harm to patient was reported but an additional graft was taken as graft was damaged.Surgery time was impacted 1-15 minutes and the patient was anesthetized.No other adverse events were reported as a result of the incident.
|
|
Manufacturer Narrative
|
This event has been recorded by zimmer biomet under (b)(4).This medwatch is being filed to relay additional information.Review of the most recent repair record determined that the motor seized during testing.The motor was replaced and resolved the reported issue.-review of the previous repair record identified unrelated repairs to the reported event.The device was tested and found to be functioning to specifications prior to release to the customer.-devices used for treatment.-review of complaint history identified additional similar complaints for the reported item and part and lot combination.Complaints are monitored through monthly complaint review in order to identify potential adverse trends.-a definitive root cause cannot be determined.-no corrective actions, preventive actions, or field actions resulted after investigation of this event.-the event is confirmed.
|
|
Event Description
|
No additional event information.
|
|
Search Alerts/Recalls
|