Model Number 3650
Device Problem Program or Algorithm Execution Failure (4036)
Patient Problem Arrhythmia (1721)
Event Date 03/15/2021
Event Type  Injury  
Manufacturer Narrative
The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.
 
Event Description
It was reported that the patient presented to the hospital for a generator replacement procedure.During capture threshold testing, the user released the button to terminate the testing as expected, however, the device continued to decrement, resulting in a loss of capture.Pacing was resumed after the test auto-terminated.The patient experienced asystole but as asymptomatic due to anesthesia.
 
Manufacturer Narrative
Correction - h6 - investigation conclusion should have been cause traced to device design rather than cause not established.
 
Manufacturer Narrative
Additional information - h7, h9 - recall number.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MERLIN¿
Type of Device
PROGRAMMER, PACEMAKER
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
sonali arangil
15900 valley view court
sylmar, CA 91342
MDR Report Key11605749
MDR Text Key243539609
Report Number2017865-2021-13511
Device Sequence Number1
Product Code NIK
UDI-Device Identifier05414734500661
UDI-Public05414734500661
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Remedial Action Recall
Type of Report Initial,Followup,Followup
Report Date 04/06/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/01/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number3650
Device Catalogue Number3650
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received03/10/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction Number2017865-03/10/22¿001¿C
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age74 YR
Patient SexMale
Patient Weight79 KG
Patient EthnicityNon Hispanic
-
-