MEDTRONIC NAVIGATION, INC (LITTLETON) O-ARM O2 IMAGING SYSTEM; INTERVENTIONAL FLUOROSCOPIC X-RAY SYSTEM
|
Back to Search Results |
|
Model Number BI70002000 |
Device Problems
Poor Quality Image (1408); Computer Operating System Problem (2898); Application Program Freezes, Becomes Nonfunctional (4031)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 03/09/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
Information references the main component of the system.Other relevant device(s) are: product id: bi71000194, serial/lot #: none, ubd: unknown, udi#: unknown.A medtronic representative visited the site to evaluate the equipment.It was found that the handswitch 2d button was intermittently working and the 3d button was not working at all.Codes b01, c07, and d02 are applicable.No other products have been returned to medtronic for analysis.Codes b17, c20, and d15 are applicable.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Medtronic received information regarding an imaging system being used outside of a procedure.It was reported that the site lags while taking 2d fluoro and cannot get the system to take a 3d spin.There was no patient involved.
|
|
Manufacturer Narrative
|
D/h2/h3/h6: the handswitch was returned and analyzed.The issue was confirmed.The handswitch was installed in a test system.The system booted and readied.The 2d and store buttons functioned correctly but they were unable to acquire 3d images using the handswitch.Codes b01, c07 and d02 are applicable.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|