• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS MEDICAL SYSTEMS CORP. HIGH FLOW INSUFFLATION UNIT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS MEDICAL SYSTEMS CORP. HIGH FLOW INSUFFLATION UNIT Back to Search Results
Model Number UHI-3
Device Problem Device Sensing Problem (2917)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
Additional information is not yet available for this event.Event date is not known.Supplemental report(s) will be filed as any information becomes available.The device has been returned and a device evaluation completed for it.Manufacture date is not known.For testing, the uhi-3 high flow insufflation unit was powered on without co2 gas inside.The pressure indicator displayed number 1 instead of number 0, and the 3 green leds of the bar graph pressure indicator lit up at the same time.This issue of the pressure gauge is caused by faulty main board.Additionally, the front panel is broken at the top left corner.The front chassis is deformed.
 
Event Description
As reported for this event, during preparation for use, the device was reading zero pressure.There is no patient involvement and no harm reported to any patient.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the legal manufacturer's investigation, it is likely the phenomenon occurred due to a failure of the pressure sensor mounted on the main board.The cause of the failure of the pressure sensor is likely to be age deterioration over time, as it has been more than 13 years since the device was manufactured.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HIGH FLOW INSUFFLATION UNIT
Type of Device
HIGH FLOW INSUFFLATION UNIT
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
MDR Report Key11608533
MDR Text Key246108768
Report Number8010047-2021-04519
Device Sequence Number1
Product Code HIF
UDI-Device Identifier04953170140280
UDI-Public04953170140280
Combination Product (y/n)N
PMA/PMN Number
K014166
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Type of Report Initial,Followup
Report Date 06/07/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberUHI-3
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/29/2021
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 01/20/2021
Initial Date FDA Received04/02/2021
Supplement Dates Manufacturer Received05/28/2021
Supplement Dates FDA Received06/07/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-