Model Number T173F6
Device Problem Material Rupture (1546)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/08/2021
Event Type  malfunction  
Manufacturer Narrative
One catheter with attached monoject 0.8 cc limited volume syringe was returned for evaluation.A non-edwards three-way stopcock was attached at the proximal injectate hub.Balloon was found to be ruptured at the central area of the balloon and the ruptured edges of balloon latex were not able to match up.All through lumens were patent without any leakage or occlusion.No visible damage or inconsistency to the catheter body, balloon bonding sites or returned syringe was observed.Visual examination was performed under microscope at 20x magnification and unaided eye.A device history record review was completed and documented that device met all specifications upon distribution.Customer report of balloon inflation issue was confirmed during the evaluation.An investigation has been initiated to consider any potential manufacturing factors that may have contributed to this complaint and implement any necessary corrective actions.Invasive procedures involve some patient risks.Although serious complications are relatively uncommon, the physician is advised, before deciding to insert or use the catheter, to consider the potential benefits in relation to the possible complications.The techniques for insertion, methods of using the catheter to obtain patient data information, and the occurrence of complications is well described in the literature.It is common clinical practice to check balloon integrity by inflating it to the recommended volume in order to detect any asymmetry or leakage condition before use of the catheter.It is unknown if user or procedural factors may have contributed to the event.Complaint histories for all reported events are reviewed against trending control limits on a monthly basis, and any excursions above the control limits are assessed and documented as part of this monthly review.
 
Event Description
It was reported that the balloon of the swan ganz catheter did not inflate after the catheter was inserted into the patient body on the first day of use.Balloon test was not performed prior to use.Patient demographic information requested but unavailable.There were no patient complications reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SWAN-GANZ POLYMER BLEND TORQUE SUPPORT TRUE SIZE T-TIP TD CATHETER
Type of Device
SWAN-GANZ CATHETER
Manufacturer (Section D)
EDWARDS LIFESCIENCES, PR
state rd indus pk 402 km 1.4
anasco PR 00610
Manufacturer (Section G)
EDWARDS LIFESCIENCES, PR
state rd indus pk 402 km 1.4
anasco PR 00610
Manufacturer Contact
jessica atallah
1 edwards way
irvine, CA 92614
9492500249
MDR Report Key11613925
MDR Text Key245233021
Report Number2015691-2021-02258
Device Sequence Number1
Product Code DQE
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K915726
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 04/05/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/05/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/09/2022
Device Model NumberT173F6
Device Catalogue NumberT173F6
Device Lot Number63397752
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/15/2021
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received03/16/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/12/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-