Model Number N/A |
Device Problems
Shipping Damage or Problem (1570); Battery Problem (2885); Expulsion (2933)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 03/15/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
This event has been recorded by zimmer biomet under (b)(4).Once the investigation is completed, a follow-up/final report will be submitted.
|
|
Event Description
|
It was reported that before shipping to the hospital, our distributor found a lot of black granular sand in the package.There was no adverse event reported as a result of this malfunction.
|
|
Manufacturer Narrative
|
This event has been recorded by zimmer biomet under (b)(4).D4: udi#: (b)(4).-visual inspection of the returned product confirmed there was an expulsion of the battery pack -review of the device history record identified no deviations or anomalies during manufacturing related to the reported event.-device is used for treatment.- complaints are monitored through monthly complaint review in order to identify potential adverse trends.-the root cause of the reported issue is attributed to a manufacturing operator error.Actions were initiated as a result of the investigation of this event.-the event is confirmed.
|
|
Event Description
|
No additional information.
|
|