CAREFUSION SD ALARIS PCA MODULE; PUMP, INFUSION Back to Search Results
Model Number 8120
Device Problems Break (1069); Pumping Stopped (1503)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
The affected device has been received and the investigation is pending.A follow up report will be submitted once the failure investigation has been completed.
 
Event Description
The customer reported broken/damaged.There was no patient involvement.
 
Event Description
The customer reported broken/damaged.There was no patient involvement.
 
Manufacturer Narrative
This device was evaluated and repaired through the service repair process.Based on the findings, the probable root cause of the reported issue was due to wear of the front cover pca.A review of the device history record showed the device had a manufacture date of 19jan2012.The review was performed from the date of manufacture to the date of product release for distribution.A review of the device history record for sn (b)(6) was performed which confirmed that this device was not involved in a production failure which correlates to the customer reported issue.A review of the complaint history record was performed for the sn (b)(6) which did not confirm similar complaints with the same or related failure mode for this customer.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALARIS PCA MODULE
Type of Device
PUMP, INFUSION
Manufacturer (Section D)
CAREFUSION SD
10020 pacific mesa blvd
san diego CA 92121 4386
MDR Report Key11618680
MDR Text Key243918173
Report Number2016493-2021-502097
Device Sequence Number1
Product Code MEA
UDI-Device Identifier10885403812002
UDI-Public10885403812002
Combination Product (y/n)N
PMA/PMN Number
K043299
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Type of Report Initial,Followup
Report Date 03/05/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/06/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number8120
Device Catalogue Number8120
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/08/2021
Date Manufacturer Received05/28/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-