Model Number N/A |
Device Problem
Difficult to Insert (1316)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 03/24/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).Concomitant medical devices: 110017330 6892575 g7 bispherical shell 48c.Foreign country: (b)(6).Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
|
|
Event Description
|
It was reported that the patient underwent a hip procedure.During the procedure, the liner would not seat into the cup.The size 48 cup was removed and replaced with a size 50 cup.Sales rep indicates no additional information is available.Attempts have been made and additional information on the reported event is unavailable at this time.
|
|
Event Description
|
No further event information available at the time of this report.
|
|
Manufacturer Narrative
|
This follow-up report is being submitted to relay additional information.Complaint sample was evaluated and the reported event was confirmed.A g7 neutral e1 liner was returned and evaluated against the complaint.Visual inspection found the liner to be in decent overall condition.Indentations and scratching were observed on the rim of the liner.The barb and scallops are undamaged.A circular indentation is present on the outer radius in 1 location with light scraping near it.The remainder of the outer radius is undamaged.The device history record (dhr) was reviewed and no discrepancies relevant to the reported event were found.Root cause was unable to be determined.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|
Search Alerts/Recalls
|