• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. UNKN PATIENT POSITIONER; APPARATUS, TRACTION, NON-POWERED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. UNKN PATIENT POSITIONER; APPARATUS, TRACTION, NON-POWERED Back to Search Results
Catalog Number UNKNOWN
Device Problem Unstable (1667)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/15/2021
Event Type  malfunction  
Manufacturer Narrative
(b)(4).
 
Event Description
It was reported that during a surgery, the foot continued to slip out of the supine hip positioning system due to the size of the patient.The foot was small so when traction was applied the foot slipped and lost the space within the joint.The procedure was successfully completed without a significant delay.It is unknown how was the procedure finished.No patient injuries were reported.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
 
Manufacturer Narrative
H3,h6: the reported device, used in treatment was not returned to the designated complaint unit for independent evaluation, thus visual inspection and functional testing could not be performed.No product identification information was provided and thus a manufacturing record review and complaint history could not be conducted.A relationship, if any, between the subject device and the reported event could not be determined.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.No containment or corrective actions are recommended at this time.If the product associated with this event is returned at a future date, this investigation will be reopened for evaluation.
 
Event Description
It was reported that during a hip arthroscopy, the foot continued to slip out of the supine hip positioning system due to the size of the patient.The foot was small so when traction was applied the foot slipped and lost the space within the joint.The procedure was successfully completed without a significant delay using the same device.No patient injuries were reported.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNKN PATIENT POSITIONER
Type of Device
APPARATUS, TRACTION, NON-POWERED
Manufacturer (Section D)
SMITH & NEPHEW, INC.
150 minuteman road
andover MA 01810
MDR Report Key11627007
MDR Text Key244200834
Report Number3003604053-2021-00139
Device Sequence Number1
Product Code HST
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,o
Type of Report Initial,Followup,Followup
Report Date 06/02/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberUNKNOWN
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 03/15/2021
Initial Date FDA Received04/07/2021
Supplement Dates Manufacturer Received03/30/2021
05/07/2021
Supplement Dates FDA Received04/12/2021
06/02/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-