Model Number 27093 |
Device Problems
Device Displays Incorrect Message (2591); Pressure Problem (3012)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 03/11/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
The device was returned to an authorized resmed third party service center and an evaluation confirmed the complaint.The device was recalibrated to address this issue.The device was serviced and fully tested before it was returned to the customer.Resmed reference #: (b)(4).
|
|
Event Description
|
It was reported to resmed that an astral device displayed an error message (sf115) related to atmospheric pressure measurement.There was no patient harm or serious injury reported as a result of this incident.
|
|
Event Description
|
It was reported to resmed that an astral device displayed an error message (sf115) related to atmospheric pressure measurement.There was no patient harm or serious injury reported as a result of this incident.
|
|
Manufacturer Narrative
|
Based on all available evidence and complaint investigations of a similar nature, an investigation determined that the reported sf115 was due to an isolated component failure within the device main circuit board.Resmed's risk analysis for this failure mode concludes that the risk is acceptable.Resmed reference #: (b)(4).
|
|