• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. CORKSCREW FEM HEAD REMOVER; PROSTH, KNEE, PATE/FEMOROTIBIAL, SEMI-CONST, UNCEM, PORO, COAT, POLY/METAL/POLY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. CORKSCREW FEM HEAD REMOVER; PROSTH, KNEE, PATE/FEMOROTIBIAL, SEMI-CONST, UNCEM, PORO, COAT, POLY/METAL/POLY Back to Search Results
Catalog Number 71930020
Device Problems Material Deformation (2976); Naturally Worn (2988)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/16/2021
Event Type  malfunction  
Event Description
It was reported that the corkscrew is worn.No case or patient involved.
 
Manufacturer Narrative
H3, h6: the device, intended for use in treatment, was not returned for evaluation.Therefore, a product analysis could not be performed at this time and the reported event could not be confirmed.A review of the device history records for the listed batch did not reveal any deviation from the standard manufacturing processes.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.A review of the complaint history for the listed part revealed prior complaints for this batch number.This device is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation warranted for this complaint; however we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CORKSCREW FEM HEAD REMOVER
Type of Device
PROSTH, KNEE, PATE/FEMOROTIBIAL, SEMI-CONST, UNCEM, PORO, COAT, POLY/METAL/POLY
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key11646857
MDR Text Key244856609
Report Number1020279-2021-02858
Device Sequence Number1
Product Code MBH
UDI-Device Identifier00885556049709
UDI-Public885556049709
Combination Product (y/n)N
PMA/PMN Number
K121393
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 05/17/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/11/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number71930020
Device Lot Number14GM12327
Was Device Available for Evaluation? Yes
Date Manufacturer Received05/16/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-