It was reported that during the burring of the distal femoral resection in a navio tka procedure, a handpiece error appeared "handpiece exposure motor control failure".The handpiece was swapped out, but the error remained.They navigated the cut to keep efficiency.There was a delay of fewer than 30 minutes.The handpiece was tested post surgery with resulting standard values.No patient harm or additional complications were reported.
|
H3, h6: the navio handpiece, part number pfsr110137, serial (b)(6) and used for treatment, was returned for evaluation.A relationship between the reported event and the device could not be established.The reported problem could not be visually confirmed.A functional evaluation was performed.The reported problem was not confirmed.The evaluation followed a performance verification.The handpiece passed all tests.No additional functional non-conformances were identified.A review of manufacturing records indicate the device met all specifications upon release into distribution.A complaint history review for similar reported/confirmed complaints has identified prior events.No probable contributing factors could be attributed to this complaint.Although no further containment or corrective action is recommended or required at this time, all complaints are monitored and trended through post market surveillance activities.
|