• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MAQUET CARDIOVASCULAR LLC VASOVIEWHEMPRO VH-3500; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAQUET CARDIOVASCULAR LLC VASOVIEWHEMPRO VH-3500; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES Back to Search Results
Model Number VASOVIEWHEMPRO VH-3500
Device Problem No Flow (2991)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/08/2021
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The device has been returned to the factory and is being evaluated.A supplemental report will be submitted when the evaluation is completed.
 
Event Description
The hospital reported that during an endoscopic vein harvesting procedure, vasoview hemopro vh-3500 would not deliver co2 (it was the btt).A replacement device was used to complete the procedure.The hospital did not report any patient effects.
 
Manufacturer Narrative
(b)(4).Updated section: g4, g7, h2, h3, h6, h10.The device was returned to the factory for evaluation on 03/25/2021.An investigation was conducted on 04/13/2021.Signs of clinical use and evidence of blood was observed on the btt.A mechanical evaluation was conducted.The btt was able to be inflated.No visual defects were observed on the silicone btt.A 5cc syringe, filled with saline was attached to the co2 line on the btt and squeezed the syringe to spray the saline.There was no backflow observed in the co2 line.Based on the condition of the device, the reported failure "no flow" was not confirmed.The lot # 25155144 history record review was completed.There were no ncmrs, rework, or deviations documented for the reported lot number.Based on the dhr/lhr review results, it was determined that there is no relation between the batch manufacturing process and the reported failure.
 
Event Description
The hospital reported that during an endoscopic vein harvesting procedure, vasoview hemopro vh-3500 would not deliver co2 (it was the btt).A replacement device was used to complete the procedure.The hospital did not report any patient effects.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VASOVIEWHEMPRO VH-3500
Type of Device
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
MAQUET CARDIOVASCULAR LLC
45 barbour pond drive
wayne NJ
MDR Report Key11648931
MDR Text Key244905625
Report Number2242352-2021-00273
Device Sequence Number1
Product Code GEI
UDI-Device Identifier00607567701250
UDI-Public00607567701250
Combination Product (y/n)N
PMA/PMN Number
K153194
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 04/12/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/12/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/30/2021
Device Model NumberVASOVIEWHEMPRO VH-3500
Device Catalogue NumberC-VH-3500
Device Lot Number25155144
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/25/2021
Date Manufacturer Received04/20/2021
Patient Sequence Number1
Patient Age50 YR
-
-