Information was received from a manufacture representative (rep) regarding a generator.It was reported that the unit failed test of the monopolar power output.The variable cut levels did not change output power (all levels oscillate about 30w, except the level 9 that was 50w out.It was noted that the variation of coagulation levels from 1 to 6 presented output about 12w, 7.8 and 10 levels were out of tolerance (80, 51 and 31w respectively).The equipment had physical damaged in the cabinet, under the energy and communication input connections had two welding points broken.The damage caused by the rigidity of the transport case (pressure exercised on the earth pin).There was no patient involved.
|