Model Number 773662
Device Problems Fluid/Blood Leak (1250); Detachment of Device or Device Component (2907)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/05/2021
Event Type  malfunction  
Manufacturer Narrative
The incident sample has been requested but to date has not been received for evaluation.If the sample is received, or if additional information pertinent to the incident is obtained a follow-up report will be submitted.As part of our manufacturing process, all device history records are reviewed and approved by quality, prior to release of product.
 
Event Description
The consumer called in to report that the feeding sets do not stay connected with his feeding tube.The feeding set disconnects causing leaking.There was no harm to the patient.
 
Manufacturer Narrative
The customers wife reported he husbands feeding sets do not stay connected with the feeding tube.There was no patient injury/harm reported.The device history record (dhr) review of the reported lot number shows evidence that the product was released according to all established procedures and qa documentation.A trend review of the reported lot number and failure mode performed did not identify any related trends.One (1) used device was returned for evaluation.Visual inspection performed determined the sample was incomplete/inadequate for evaluation; therefore, functional evaluation was unable to be performed.The reported issue was unable to be confirmed and a root cause was not determined.Additional action will not be taken at this time.This complaint will be used for tracking and trending purposes.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EPUMP FEED AND 1000MLFLUSH SET
Type of Device
PUMP, INFUSION, ENTERAL
Manufacturer (Section D)
COVIDIEN
15 hampshire street
mansfield MA 02048
Manufacturer (Section G)
COVIDIEN
15 hampshire street
mansfield MA 02048
Manufacturer Contact
jill saraiva
15 hampshire street
mansfield, MA 02048
5086183640
MDR Report Key11656672
MDR Text Key245136435
Report Number1282497-2021-10049
Device Sequence Number1
Product Code LZH
UDI-Device Identifier30884521022106
UDI-Public30884521022106
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,consum
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 06/28/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/13/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model Number773662
Device Catalogue Number773662
Device Lot Number210180050
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Date Manufacturer Received04/05/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/31/2021
Type of Device Usage N