• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION SENTINEL CEREBRAL PROTECTION SYSTEM (US); EMBOLIC PROTECTION DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION SENTINEL CEREBRAL PROTECTION SYSTEM (US); EMBOLIC PROTECTION DEVICE Back to Search Results
Model Number CMS15-10C-US
Device Problem Positioning Problem (3009)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/16/2021
Event Type  malfunction  
Event Description
Protected tavr study.It was reported that the distal filter of the sentinel cerebral protection system failed to maintain position in the carotid artery.A sentinel cerebral protection system was selected for use in a patient with minimal vascular tortuosity and mild calcification.Prior to the index procedure, heparin or another anticoagulant was given.The subject was on a prior regimen of aspirin and antiplatelets other than aspirin at the time of index procedure.The subject received a loading dose of aspirin.An introducer sheath was placed into the radial artery and the sentinel system was inserted with the proximal filter into brachiocephalic artery, and distal filter into left common carotid artery.During pre-balloon aortic valvuloplasty(bav) the distal filter was displaced from left carotid artery into the aorta.The procedure was continued and there was correct positioning of a non-bsc valve in the correct anatomical location.Post dilatation (bav) was performed and the procedure was completed.The sentinel cerebral protection system was successfully retrieved without any complications.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SENTINEL CEREBRAL PROTECTION SYSTEM (US)
Type of Device
EMBOLIC PROTECTION DEVICE
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
160 knowles drive
los gatos CA 95032
Manufacturer Contact
jay johnson
two scimed place
maple grove, MN 55311
7634942574
MDR Report Key11657432
MDR Text Key245183329
Report Number2134265-2021-04777
Device Sequence Number1
Product Code PUM
UDI-Device Identifier00863229000004
UDI-Public00863229000004
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K192460
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 04/13/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/13/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/10/2022
Device Model NumberCMS15-10C-US
Device Catalogue NumberCMS15-10C-US
Device Lot Number0025997745
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/17/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/10/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age66 YR
-
-