• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. SUPINE TABLE ATTACHMENT (A2) W/PAD; APPARATUS, TRACTION, NON-POWERED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. SUPINE TABLE ATTACHMENT (A2) W/PAD; APPARATUS, TRACTION, NON-POWERED Back to Search Results
Model Number 72201725
Device Problem Unstable (1667)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/23/2021
Event Type  malfunction  
Manufacturer Narrative
Internal complaint reference (b)(4).
 
Event Description
It was reported that the supine table attachment did not lock away properly during set up for a hip scope.The screw was striped so the attachment could not be securely attached to the bed.There was a smith and nephew back up device.No patient injury, delays or other complications were reported.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
 
Manufacturer Narrative
H3,h6: the reported device, intended for use in treatment was not returned to the designated complaint unit for independent evaluation, thus visual inspection and functional testing could not be performed.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specification upon release for distribution.A complaint history review concluded this was a repeat issue.A relationship, if any, between the subject device and the reported event could not be determined.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.If the product associated with this event is returned at a future date, this investigation will be reopened for evaluation.
 
Manufacturer Narrative
H10: internal complaint reference: (b)(4).The reported device, intended for use in treatment, was received for evaluation.There was a relationship found between the returned device and the reported incident.A visual inspection found rust on the sliding housings attached to the bed rails.It was also found that one of the two mast holder knobs was bent.A functional evaluation found one of the two mast holder knobs would not tighten due to the bend in the shaft.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specification upon release for distribution.A complaint history review concluded this was a repeat issue.The complaint was verified.Factors that could have contributed to this event include an impact force inconsistent with normal use.No containment or corrective actions are recommended at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SUPINE TABLE ATTACHMENT (A2) W/PAD
Type of Device
APPARATUS, TRACTION, NON-POWERED
Manufacturer (Section D)
SMITH & NEPHEW, INC.
150 minuteman road
andover MA 01810
MDR Report Key11663425
MDR Text Key245355379
Report Number3003604053-2021-00150
Device Sequence Number1
Product Code HST
UDI-Device Identifier03596010612816
UDI-Public03596010612816
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Type of Report Initial,Followup,Followup
Report Date 10/05/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number72201725
Device Catalogue Number72201725
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/11/2021
Initial Date Manufacturer Received 03/23/2021
Initial Date FDA Received04/14/2021
Supplement Dates Manufacturer Received05/13/2021
10/04/2021
Supplement Dates FDA Received05/14/2021
10/05/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-