• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EDWARDS LIFESCIENCES, PR PEDIASAT OXIMETRY CATHETER; CATHETER, OXIMETER, FIBER-OPTIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EDWARDS LIFESCIENCES, PR PEDIASAT OXIMETRY CATHETER; CATHETER, OXIMETER, FIBER-OPTIC Back to Search Results
Model Number XT358SP
Device Problem Leak/Splash (1354)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/26/2021
Event Type  malfunction  
Manufacturer Narrative
The device was not returned for evaluation; it was discarded at the hospital.Without return of the unit it is not possible to determine if some damage or defect existed on the unit that could have contributed to the event.It is not known if some procedural factors may have contributed to the event.No corrective actions will be taken at this time.Lot number was not provided; therefore, review of the manufacturing records could not be completed.An investigation has been initiated to consider any potential manufacturing factors that may have contributed to this complaint.Invasive procedures involve some patient risks.Although serious complications are relatively uncommon, the physician is advised, before deciding to insert or use the catheter, to consider the potential benefits in relation to the possible complications.The techniques for insertion, methods of using the catheter to obtain patient data information, and the occurrence of complications is well described in the literature.In this case, leakage occurred in a pediatric patient, who would have a lower threshold for brief interruptions in therapy.It is unknown whether user or procedural factors contributed to the stated event.Complaint histories for all reported events are reviewed against trending control limits on a monthly basis, and any excursions above the control limits are assessed and documented as part of this monthly review.
 
Event Description
It was reported that the leakage (not blood) occurred from the backform of the pediasat catheter.There was no allegation of patient injury.The devices were not available for evaluation as they were discarded at the hosptial.No other specific details could be provided, as the event occurred some time ago and the report was obtained during discussion of another event.
 
Manufacturer Narrative
Although this complaint was not confirmed by a non-conformance as the product was not returned for evaluation, another complaint with a similar system of interlumen leak came from the same reporter.Therefore, this complaint is part of a pra and capa to investigate interlumen leakage.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PEDIASAT OXIMETRY CATHETER
Type of Device
CATHETER, OXIMETER, FIBER-OPTIC
Manufacturer (Section D)
EDWARDS LIFESCIENCES, PR
state rd indus pk 402 km 1.4
anasco PR 00610
MDR Report Key11664090
MDR Text Key248854883
Report Number2015691-2021-02448
Device Sequence Number1
Product Code DQE
Combination Product (y/n)N
PMA/PMN Number
K053609
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,o
Type of Report Initial,Followup
Report Date 06/24/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberXT358SP
Device Catalogue NumberXT358SP
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 03/26/2021
Initial Date FDA Received04/14/2021
Supplement Dates Manufacturer Received06/21/2021
Supplement Dates FDA Received06/24/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-