Model Number G138 |
Device Problems
High impedance (1291); Noise, Audible (3273)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 03/08/2021 |
Event Type
malfunction
|
Event Description
|
It was reported that during an implant procedure of a cardiac resynchronization therapy defibrillator (crt-d) device there were observations of high out-of-range pacing impedances on the left ventricular (lv) lead measuring greater than 3000 ohms.Technical services (ts) recommended to reconnect the lead into the device header and impedances measured in the 2000 ohms range.Additionally, ts recommended to test with the pacing system analyzer (psa) however, the physician elected to leave it connected and programmed to a vector and impedances were within range.This crt-d and lv lead remains in service.No adverse patient effects were reported.
|
|
Event Description
|
It was reported that during an implant procedure of a cardiac resynchronization therapy defibrillator (crt-d) device there were observations of high out-of-range pacing impedances on the left ventricular (lv) lead measuring greater than 3000 ohms.Technical services (ts) recommended to reconnect the lead into the device header and impedances measured in the 2000 ohms range.Additionally, ts recommended to test with the pacing system analyzer (psa) however, the physician elected to leave it connected and programmed to a vector and impedances were within range.This crt-d and lv lead remains in service.No adverse patient effects were reported.
|
|
Manufacturer Narrative
|
This product has not been returned to boston scientific, and as a result, laboratory analysis could not be conducted.The associated investigation determined that this device exhibited intermittent impedance measurements with no conclusive evidence of a malfunction or inadequate lead-to-device connection; please refer to the description for more information regarding the specific circumstances of this event.Investigation has determined that this type of event is likely the result of an intermittent high impedance condition associated with the device spring contact and lead terminal ring.A design enhancement was implemented in 2020 to stabilize the electrical connection between the spring contact and the lead terminal.
|
|
Event Description
|
It was reported that during an implant procedure of a cardiac resynchronization therapy defibrillator (crt-d) device there were observations of high out-of-range pacing impedances on the left ventricular (lv) lead measuring greater than 3000 ohms.Technical services (ts) recommended to reconnect the lead into the device header and impedances measured in the 2000 ohms range.Additionally, ts recommended to test with the pacing system analyzer (psa) however, the physician elected to leave it connected and programmed to a vector and impedances were within range.This crt-d and lv lead remains in service.No adverse patient effects were reported.It was determined that the high out-of-range pacing impedances on the lv lead was not associated with the device's spring contact and lead terminal ring as previously reported.As all available evidence indicates the cause of the event was due to the lv lead and not the device, this event is no longer deemed to be a reportable incident.
|
|
Search Alerts/Recalls
|