The returned device was thoroughly inspected and analyzed upon receipt at our post market quality assurance laboratory.The lv set screw was found stuck in the returned non-boston scientific hex wrench.Pin gauge testing, designed to verify proper port dimensions, was completed.Each port measured as expected.The device was then exposed to simulated heart load conditions, and the pacing, and sensing functions were tested.Impedance testing was completed and all measurements were within normal limits.The device operated appropriately with no interruptions in therapy output at the returned programmed settings.A series of electrical tests was also performed, and again, normal device function was observed.It was concluded that the use of a non-boston scientific hex wrench contributed to the clinical observations.
|