Received one 000150 in opened original packaging.Lot number was verified.Performed a visual inspection, the guidewire has a clean break at the spring.All pieces were returned.The manufacturing documents from the device history record have been reviewed with special attention to the manufacturing and inspection of the product.The product released for distribution was found to have met all specifications prior to shipment.This is the only complaint for this lot number and failure mode within the past two years.A two-year review of complaint history revealed there has been a total of 8 complaints, regarding 17 devices, for this device family and failure mode.During this same time frame (b)(4) devices have been manufactured and shipped worldwide.Should all the complaint devices have been found confirmed for this reported failure, the rate of failure would be (b)(4).Per the instructions for use, the user is advised the following: the guidewire should be not advanced if resistance is met without determining the cause and taking remedial action.Before and after each use, carefully inspect the guidewire for wear, damage or abnormal bending.The entire wire should be inspected in this manner, but areas of extra focus include the flexible spring tip and the soldered joints between the spring tip and the wire.If the joints appear discolored, loose or cracked, discard the guidewire.If wear, damage or abnormal bending is found at any location on the guidewire discard the guidewire.This issue will continue to be monitored through the complaint system to assure patient safety.
|